Method development of stir bar sportive extraction coupled with thermal desorption-gas chromatography-mass spectrometry for the analysis of phthalates in Peruvian pisco

Gebrehiwot D.G.; Castro R.; Hidalgo-Gárate J.C.; Robles A.D.; Durán-Guerrero E. · Journal of Chromatography A (2023) · DOI: 10.1016/j.chroma.2023.464470

Resumen

In this work, for the first time, a stir bar sorptive extraction (SBSE) coupled with thermal desorption–gas chromatography–mass spectrometry (TD–GC–MS) was developed and validated for the determination of seven phthalates in Peruvian pisco. The phthalate compounds considered were dimethyl phthalate (DMP), diethyl phthalate (DEP), bis(2-ethylhexyl) hexahydrophthalate (BEHP), benzyl butyl phthalate (BBP), di-n-butyl phthalate (DBP), di-isodecyl phthalate (DIDP) and di-isobutyl phthalate (DIBP). The best overall analytical conditions obtained from the optimization were as follow: extraction time of 120 min, size of polydimethylsiloxane (PDMS) twister (20 mm length x 1 mm thickness), NaCl content (20 %) and sample volume (40 mL). The in-house validation of SBSE/TD–GC–MS method was performed taking into account the ISO/IEC 17,025 requirements and EURACHEM/CITAC guideline. Under optimal conditions, very low limits of detection of 1.3–0.21 µg L−1 were obtained. Furthermore, the limits of quantification ranged from 4.2–70 µg L−1, and the correlation coefficients were found to be ≥ 0.991. The method was precise, with relative standard deviations (RSD, %) for inter twister repeatability and the inter day repeatability precisions from 1.1 to 11 and from 6.2 to 15.9, respectively. The pisco samples were analysed with recoveries between 91–124.4%, and DBP, BEHP, and BBP were the most commonly found compounds in the samples. The optimized methodology was also evaluated in terms of green character, and it obtained almost the best AGREE score when it was compared with other previous methods for the analysis of phthalates in alcoholic beverages. Therefore, the SBSE/TD–GC–MS method has proved to be suitable for routine practice because it is simple, less laborious, economical, precise, accurate and green, and it would be applicable for pisco safety regulations.

Instituciones: Universidad Privada San Juan Bautista